From Wikipedia, the free encyclopedia
 
		  Beckman DU640 UV/Vis spectrophotometer
 
UV spectroscopy or UV–visible spectrophotometry (UV–Vis or UV/Vis) refers to absorption spectroscopy or reflectance spectroscopy in part of the ultraviolet and the full, adjacent visible regions of the electromagnetic spectrum.
 Being relatively inexpensive and easily implemented, this methodology 
is widely used in diverse applied and fundamental applications.  The 
only requirement is that the sample absorb in the UV-vis region, i.e. be
 a chromophore. Absorption spectroscopy is complementary to fluorescence spectroscopy.
  Parameters of interest, besides the wavelength of measurement, are 
absorbance (A) or Transmittance (%T) or Reflectance (%R), and its change
 with time.
Optical transitions
Most
 molecules and ions absorb energy in the ultraviolet or visible range, 
i.e., they are chromophores.  The absorbed photon excites an electron in
 the chromophore to higher energy molecular orbitals, giving rise to an excited state.
 For organic chromophores, four possible types of transitions are 
assumed: π–π*, n–π*, σ–σ*, and n–σ*.  Transition metal complexes are 
often colored (i.e., absorb visible light) owing to the presence of 
multiple electronic states associated with incompletely filled d 
orbitals.
Applications
An example of a UV/Vis readout
 
UV/Vis spectroscopy is routinely used in analytical chemistry for the quantitative determination of diverse analytes or sample, such as transition metal ions, highly conjugated organic compounds,
 and biological macromolecules. Spectroscopic analysis is commonly 
carried out in solutions but solids and gases may also be studied.
- Organic compounds, especially those with a high degree of conjugation, also absorb light in the UV or visible regions of the electromagnetic spectrum. The solvents for these determinations are often water for water-soluble compounds, or ethanol
 for organic-soluble compounds. (Organic solvents may have significant 
UV absorption; not all solvents are suitable for use in UV spectroscopy.
 Ethanol absorbs very weakly at most wavelengths.) Solvent polarity and 
pH can affect the absorption spectrum of an organic compound. Tyrosine, 
for example, increases in absorption maxima and molar extinction 
coefficient when pH increases from 6 to 13 or when solvent polarity 
decreases.
 - While charge transfer complexes also give rise to colours, the colours are often too intense to be used for quantitative measurement.
 
The Beer–Lambert law
 states that the absorbance of a solution is directly proportional to 
the concentration of the absorbing species in the solution and the path 
length.
 Thus, for a fixed path length, UV/Vis spectroscopy can be used to 
determine the concentration of the absorber in a solution. It is 
necessary to know how quickly the absorbance changes with concentration.
 This can be taken from references (tables of molar extinction coefficients), or more accurately, determined from a calibration curve.
A UV/Vis spectrophotometer may be used as a detector for HPLC.
 The presence of an analyte gives a response assumed to be proportional 
to the concentration. For accurate results, the instrument's response to
 the analyte in the unknown should be compared with the response to a 
standard; this is very similar to the use of calibration curves. The 
response (e.g., peak height) for a particular concentration is known as 
the response factor.
The wavelengths of absorption peaks can be correlated with the 
types of bonds in a given molecule and are valuable in determining the 
functional groups within a molecule. The Woodward–Fieser rules, for instance, are a set of empirical observations used to predict λmax, the wavelength of the most intense UV/Vis absorption, for conjugated organic compounds such as dienes and ketones.
 The spectrum alone is not, however, a specific test for any given 
sample. The nature of the solvent, the pH of the solution, temperature, 
high electrolyte concentrations, and the presence of interfering 
substances can influence the absorption spectrum. Experimental 
variations such as the slit width (effective bandwidth) of the 
spectrophotometer will also alter the spectrum. To apply UV/Vis 
spectroscopy to analysis, these variables must be controlled or 
accounted for in order to identify the substances present.
The method is most often used in a quantitative way to determine concentrations of an absorbing species in solution, using the Beer–Lambert law:
,
where A is the measured absorbance (in Absorbance Units (AU)), 
 is the intensity of the incident light at a given wavelength, 
 is the transmitted intensity, L the path length through the sample, and c the concentration of the absorbing species. For each species and wavelength, ε is a constant known as the molar absorptivity
 or extinction coefficient. This constant is a fundamental molecular 
property in a given solvent, at a particular temperature and pressure, 
and has units of 
.
The absorbance and extinction ε are sometimes defined in terms of the natural logarithm instead of the base-10 logarithm.
The Beer–Lambert Law is useful for characterizing many compounds 
but does not hold as a universal relationship for the concentration and 
absorption of all substances. A 2nd order polynomial relationship 
between absorption and concentration is sometimes encountered for very 
large, complex molecules such as organic dyes (Xylenol Orange or Neutral Red, for example).
UV–Vis spectroscopy is also used in the semiconductor industry to
 measure the thickness and optical properties of thin films on a wafer. 
UV–Vis spectrometers are used to measure the reflectance of light, and 
can be analyzed via the Forouhi–Bloomer dispersion equations to determine the index of refraction (
) and the extinction coefficient (
) of a given film across the measured spectral range.
Practical considerations
The
 Beer–Lambert law has implicit assumptions that must be met 
experimentally for it to apply; otherwise there is a possibility of 
deviations from the law.
 For instance, the chemical makeup and physical environment of the 
sample can alter its extinction coefficient. The chemical and physical 
conditions of a test sample therefore must match reference measurements 
for conclusions to be valid. Worldwide, pharmacopoeias such as the 
American (USP) and European (Ph. Eur.) pharmacopeias demand that 
spectrophotometers perform according to strict regulatory requirements 
encompassing factors such as stray light and wavelength accuracy. 
Spectral bandwidth
It is important to have a monochromatic source of radiation for the light incident on the sample cell.
 Monochromaticity is measured as the width of the "triangle" formed by 
the intensity spike, at one half of the peak intensity. A given 
spectrometer has a spectral bandwidth that characterizes how monochromatic the incident light is. If this bandwidth is comparable to (or more than) the width
 of the absorption line, then the measured extinction coefficient will 
be mistaken. In reference measurements, the instrument bandwidth 
(bandwidth of the incident light) is kept below the width of the 
spectral lines. When a test material is being measured, the bandwidth of
 the incident light should also be sufficiently narrow. Reducing the 
spectral bandwidth reduces the energy passed to the detector and will, 
therefore, require a longer measurement time to achieve the same signal 
to noise ratio.
Wavelength error
In
 liquids, the extinction coefficient usually changes slowly with 
wavelength. A peak of the absorbance curve (a wavelength where the 
absorbance reaches a maximum) is where the rate of change in absorbance 
with wavelength is smallest.
 Measurements are usually made at a peak to minimize errors produced by 
errors in wavelength in the instrument, that is errors due to having a 
different extinction coefficient than assumed.
Stray light
Another important major factor is the purity of the light used. The most important factor affecting this is the stray light level of the monochromator.
The detector used is broadband; it responds to all the light that
 reaches it. If a significant amount of the light passed through the 
sample contains wavelengths that have much lower extinction coefficients
 than the nominal one, the instrument will report an incorrectly low 
absorbance. Any instrument will reach a point where an increase in 
sample concentration will not result in an increase in the reported 
absorbance, because the detector is simply responding to the stray 
light. In practice the concentration of the sample or the optical path 
length must be adjusted to place the unknown absorbance within a range 
that is valid for the instrument. Sometimes an empirical calibration 
function is developed, using known concentrations of the sample, to 
allow measurements into the region where the instrument is becoming 
non-linear.
As a rough guide, an instrument with a single monochromator would
 typically have a stray light level corresponding to about 3 Absorbance 
Units (AU), which would make measurements above about 2 AU problematic. A
 more complex instrument with a double monochromator
 would have a stray light level corresponding to about 6 AU, which would
 therefore allow measuring a much wider absorbance range.
Deviations from the Beer–Lambert law
At
 sufficiently high concentrations, the absorption bands will saturate 
and show absorption flattening. The absorption peak appears to flatten 
because close to 100% of the light is already being absorbed. The 
concentration at which this occurs depends on the particular compound 
being measured. One test that can be used to test for this effect is to 
vary the path length of the measurement. In the Beer–Lambert law, 
varying concentration and path length has an equivalent effect—diluting a
 solution by a factor of 10 has the same effect as shortening the path 
length by a factor of 10. If cells of different path lengths are 
available, testing if this relationship holds true is one way to judge 
if absorption flattening is occurring.
Solutions that are not homogeneous can show deviations from the 
Beer–Lambert law because of the phenomenon of absorption flattening. 
This can happen, for instance, where the absorbing substance is located 
within suspended particles.
 The deviations will be most noticeable under conditions of low 
concentration and high absorbance. The last reference describes a way to
 correct for this deviation.
Some solutions, like copper(II)chloride in water, change visually
 at a certain concentration because of changed conditions around the 
coloured ion (the divalent copper ion). For copper(II)chloride it means a
 shift from blue to green, which would mean that monochromatic measurements would deviate from the Beer–Lambert law.
Measurement uncertainty sources
The above factors contribute to the measurement uncertainty
 of the results obtained with UV/Vis spectrophotometry. If UV/Vis 
spectrophotometry is used in quantitative chemical analysis then the 
results are additionally affected by uncertainty sources arising from 
the nature of the compounds and/or solutions that are measured. These 
include spectral interferences caused by absorption band overlap, fading
 of the color of the absorbing species (caused by decomposition or 
reaction) and possible composition mismatch between the sample and the 
calibration solution.
Ultraviolet–visible spectrophotometer
The instrument used in ultraviolet–visible spectroscopy is called a UV/Vis spectrophotometer. It measures the intensity of light after passing through a sample (
), and compares it to the intensity of light before it passes through the sample (
). The ratio 
 is called the transmittance, and is usually expressed as a percentage (%T). The absorbance, 
, is based on the transmittance:

The UV–visible spectrophotometer can also be configured to measure 
reflectance. In this case, the spectrophotometer measures the intensity 
of light reflected from a sample (
), and compares it to the intensity of light reflected from a reference material (
) (such as a white tile). The ratio 
 is called the reflectance, and is usually expressed as a percentage (%R).
The basic parts of a spectrophotometer are a light source, a holder for the sample, a diffraction grating in a monochromator or a prism to separate the different wavelengths of light, and a detector. The radiation source is often a Tungsten filament (300–2500 nm), a deuterium arc lamp, which is continuous over the ultraviolet region (190–400 nm), Xenon arc lamp, which is continuous from 160 to 2,000 nm; or more recently, light emitting diodes (LED) for the visible wavelengths. The detector is typically a photomultiplier tube, a photodiode, a photodiode array or a charge-coupled device
 (CCD). Single photodiode detectors and photomultiplier tubes are used 
with scanning monochromators, which filter the light so that only light 
of a single wavelength reaches the detector at one time. The scanning 
monochromator moves the diffraction grating to "step-through" each 
wavelength so that its intensity may be measured as a function of 
wavelength. Fixed monochromators are used with CCDs and photodiode 
arrays. As both of these devices consist of many detectors grouped into 
one or two dimensional arrays, they are able to collect light of 
different wavelengths on different pixels or groups of pixels 
simultaneously.
Simplified schematic of a double beam UV–visible spectrophotometer
 
A spectrophotometer can be either single beam or double beam. In a single beam instrument (such as the Spectronic 20), all of the light passes through the sample cell. 
 must be measured by removing the sample. This was the earliest design 
and is still in common use in both teaching and industrial labs.
In a double-beam instrument, the light is split into two beams 
before it reaches the sample. One beam is used as the reference; the 
other beam passes through the sample. The reference beam intensity is 
taken as 100% Transmission (or 0 Absorbance), and the measurement 
displayed is the ratio of the two beam intensities. Some double-beam 
instruments have two detectors (photodiodes), and the sample and 
reference beam are measured at the same time. In other instruments, the 
two beams pass through a beam chopper,
 which blocks one beam at a time. The detector alternates between 
measuring the sample beam and the reference beam in synchronism with the
 chopper. There may also be one or more dark intervals in the chopper 
cycle. In this case, the measured beam intensities may be corrected by 
subtracting the intensity measured in the dark interval before the ratio
 is taken.
In a single-beam instrument, the cuvette containing only a 
solvent has to be measured first. Mettler Toledo developed a single beam
 array spectrophotometer that allows fast and accurate measurements over
 the UV/VIS range. The light source consists of a Xenon flash lamp for 
the ultraviolet (UV) as well as for the visible (VIS) and near-infrared 
wavelength regions covering a spectral range from 190 up to 1100 nm. The
 lamp flashes are focused on a glass fiber which drives the beam of 
light onto a cuvette containing the sample solution. The beam passes 
through the sample and specific wavelengths are absorbed by the sample 
components. The remaining light is collected after the cuvette by a 
glass fiber and driven into a spectrograph. The spectrograph consists of
 a diffraction grating that separates the light into the different 
wavelengths, and a CCD sensor to record the data, respectively. The 
whole spectrum is thus simultaneously measured, allowing for fast 
recording. 
Samples for UV/Vis spectrophotometry are most often liquids, 
although the absorbance of gases and even of solids can also be 
measured. Samples are typically placed in a transparent cell, known as a cuvette. Cuvettes are typically rectangular in shape, commonly with an internal width of 1 cm. (This width becomes the path length, 
, in the Beer–Lambert law.) Test tubes
 can also be used as cuvettes in some instruments. The type of sample 
container used must allow radiation to pass over the spectral region of 
interest. The most widely applicable cuvettes are made of high quality fused silica or quartz glass
 because these are transparent throughout the UV, visible and near 
infrared regions. Glass and plastic cuvettes are also common, although 
glass and most plastics absorb in the UV, which limits their usefulness 
to visible wavelengths.
Specialized instruments have also been made. These include 
attaching spectrophotometers to telescopes to measure the spectra of 
astronomical features. UV–visible microspectrophotometers consist of a 
UV–visible microscope integrated with a UV–visible spectrophotometer.
A complete spectrum of the absorption at all wavelengths of 
interest can often be produced directly by a more sophisticated 
spectrophotometer. In simpler instruments the absorption is determined 
one wavelength at a time and then compiled into a spectrum by the 
operator. By removing the concentration dependence, the extinction 
coefficient (ε) can be determined as a function of wavelength.
Microspectrophotometry
UV–visible
 spectroscopy of microscopic samples is done by integrating an optical 
microscope with UV–visible optics, white light sources, a monochromator, and a sensitive detector such as a charge-coupled device (CCD) or photomultiplier
 tube (PMT). As only a single optical path is available, these are 
single beam instruments. Modern instruments are capable of measuring 
UV–visible spectra in both reflectance and transmission of micron-scale 
sampling areas. The advantages of using such instruments is that they 
are able to measure microscopic samples but are also able to measure the
 spectra of larger samples with high spatial resolution. As such, they 
are used in the forensic laboratory to analyze the dyes and pigments in 
individual textile fibers, microscopic paint chips and the color of glass fragments. They are also used in materials 
science and biological research and for determining the energy content 
of coal and petroleum source rock by measuring the vitrinite
 reflectance. Microspectrophotometers are used in the semiconductor and 
micro-optics industries for monitoring the thickness of thin films after
 they have been deposited. In the semiconductor industry, they are used 
because the critical dimensions of circuitry is microscopic. A typical 
test of a semiconductor wafer would entail the acquisition of spectra 
from many points on a patterned or unpatterned wafer. The thickness of 
the deposited films may be calculated from the interference pattern
 of the spectra. In addition, ultraviolet–visible spectrophotometry can 
be used to determine the thickness, along with the refractive index and 
extinction coefficient of thin films. A map of the film thickness across the entire wafer can then be generated and used for quality control purposes.
Additional applications
UV/Vis can be applied to characterize the rate of a chemical reaction.
 Illustrative is the conversion of the yellow-orange and blue isomers of
 mercury dithizonate. This method of analysis relies on the fact that 
concentration is linearly proportional to concentration.  In the same 
approach allows determination of equilibria between chromophores. 
From the spectrum of burning gases, it is possible to determine a
 chemical composition of a fuel, temperature of gases, and air-fuel 
ratio.